850G-1040 Fluoropolymer Coating

850G-1040 Water-Based Surface Chemistry Technology

Next-Generation Surface Chemistry

Miller-Stephenson Medical in unison with Chemours™ has developed a series of water-based, medical coatings which offer next-generation performance, surface adhesion, durability, and superior surface smoothness all while eliminating the use of chromium VI. Its lower hazardous classification and stability at ambient temperature also help with handling, application pot-life, and storage conditions. The 850G-104X not only provides our smoothest surface finish but does so with an average film thickness of only 6-12 microns. 

Coating Uses

  • Hospital and Lab Equipment
  • Dental Equipment
  • Surgical Instruments
  • Electrosurgical Devices
  • Ask our team if your device or component can be coated!

Coating Equipment & Application Recommendations

All of our PTFE coatings can be applied via spray, dip or spool-to-spool methods. Proper pre-coating wire cleaning treatment and post-application curing equipment is critical to guarantee performance and durability. Acceptable curing equipment includes high-temperature ovens which can safely maintain temperatures of 246 °C  (475 °F) or utilize IR ovens. This will ensure proper sintering of the coatings onto the wire and promote the best surface characteristics that can be achieved.

Coating Property Data

ColorGreen
Typical Film Thickness6 – 12 microns
Coverage,2 m2/kg (m2/L) (ft2/gal)4.44 (4.88) (199)
Viscosity,3 centipoises400 – 600
Volume Solids, %10.2 – 14.2
Weight Solids,4 %18.0 – 21.0
Density, kg/l (lbs/gal)1.10 (9.17)
VOC content, U.S. lb/gal (g/L) (less exempt)
As Packaged0.401 (48.0)
Less Exempt3.93 (472)
Maximum In-Use Temperature, °C (°F)260 (500)
Flash Point, SETA closed cup, °C (°F)>93 (200)

ASTM & ISO Standards Met

 

  • ASTM 1894 Coefficient of Friction
  • ASTM D3359 Adhesion Test
  • ISO 7802 Wire Wrap Test
  • ISO 11070 Corrosion Resistance Test
  • Coated smooth stainless steel panels passed Adhesion test, ASTM-3359
  • Coated stainless steel catheter guidewires passed flexibility/adhesion tests, ISO 7802 Annex F wrapping test
  • Coated stainless steel catheter guidewires passed corrosion resistance, ISO11070 Annex B, saline soak test

All Miller-Stephenson’s formulations generate very uniform thin films which exceed all ISO and ASTM requirements.

Do you currently work with a PTFE coating that isnt meeting your requirements?

Looking to qualify or replace an existing PTFE coating?

With a wide array of coatings, you can qualify a new PTFE coating here. 

Don't have the coating equipment or experience?

This exceptional coating service is offered to any customer or OEM that does not have the ability to coat in-house or is looking to have a complete pre-coated turn-key solution product made and delivered. We offer customers the ability to utilize the coating services of the most highly regarded medical coaters in the United States. They provide strictly medical coating operations on some of the world’s most difficult geometrical device designs.